Rubber/Plastic Test Machines

UP-5025 Plastic Film Thickness Tester ISO 4593 ISO 534 ISO 3034

UP-5025 Plastic Film Thickness Tester ISO 4593 ISO 534 ISO 3034(图1)

Plastic Film Thickness Tester ISO 4593 ISO 534 ISO 3034

Use

Thickness Tester is designed based on the mechanical contacting method, which effectively ensures standard and accurate test data and applicable to thickness test of plastic films, sheets, diaphragms, paper, foils, silicon wafers and other materials within specified range.

Character

Contact area and pressure are strictly designed according to standard requirements, while customization is also available

Automatic lifting presser foot facilitates to minimize system errors that caused by human factors during the test

Manual or automatic operating mode for convenient test

Automatic specimen feeding, specimen feeding interval, number of testing points and specimen feeding speed could be preset by the user.

Displays the real-time data of maximum, minimum, average and standard deviation value for data analysis

Automatic statistics and printing functions are available which is convenient for the user to obtain the test results

Equipped with standard block for system calibration to ensure uniform and accurate test data

The instrument is controlled by micro-computer with LCD display, PVC operation panel and menu interface

Equipped with RS232 port which is convenient for data transfer

Test standard

ISO 4593,ISO 534,ISO 3034,GB/T 6672,GB/T 451.3, GB/T 6547,ASTM D374,ASTM D1777,TAPPI T411,JIS K6250,JIS K6783,JIS Z1702, BS 3983,BS 4817

Applications Specifications

Basic   Applications

Plastic Films, Sheets and Diaphragms

Paper and Paper Board

Foils and Silicon Wafers

Metal Sheets

Textiles and Non-woven Fabrics, e.g. baby   diapers, sanitary towel and other sheeting

Solid Electrical Insulating Materials

 

Extended   Applications

Extended Test Range of 5mm and 10mm

Curved Presser Foot

 

Speicifcation

Test   Range

0~2 mm (standard)
  0~6 mm, 12 mm (optional)

Resolution

0.1 μm

Test Speed

10 times/min (adjustable)

Test Pressure

17.5±1 KPa (film)
  50±1 KPa (paper)

Contact Area

50 mm2 (film)
  200 mm2 (paper)
  Note: Select one presser foot for film or paper; Customization is available

Specimen Feeding Interval

0 ~ 1000 mm

Specimen Feeding Speed

0.1 ~ 99.9 mm/s

Instrument Dimension

461 mm (L) x 334 mm (W) x 357 mm (H)

Power Supply

AC 220V 50Hz

Net Weight

32 kg

 

Standard configuration
A Standard gauge block, professional l software, communication cable, measuring head